Streckeisen

Community Broadband Sensor Repair Policy

ver. 2011093

The PASSCAL Instrument Center (PIC) has historically honored requests to repair community owned broadband seismometers. Significant resources are required to process, evaluate, and repair a broadband instrument. To ensure that we are able to properly administer these requests and that we are offering these services consistently to all members of the IRIS community, the following policy will govern the PASSCAL Instrument Center's handling of these requests.

Streckeisen STS-2 Broadband Sensor

Salient Features: Flat response to velocity from 120 seconds to about 10 Hz Low power Installation Tips: Example of STS-2/RT130 Setup One-pager field sheet

  Troubleshooting Tips: If you cannot center the masses of even one of the 3 elements of the STS-2, all your data could be compromised. The document provided here will guide you to a 'manual' process for attempting to get those masses centered, IF and ONLY IF the masses do not center after repeated (a dozen tries) centering commands.  